Thermal conditioned cabinet for TGR (for Europe 220/240 Volt EU plug)

Product Code:35680-EU

An essential optional tool for Your TGR experiments. This cabinet is made to efficiently control the inside temperature from 0° up to 70° and can hold up to two TGR apparatus.

In TGR experiments temperature is critical; to perform the test (which normally takes 1 hour) You need to ensure the TGR apparatus is placed into an environment having the average temperature between the two temperatures set on the TGR machine. Typical TGR set-up temps are:

  • Cold 5°C and Hot 30°C: average temperature is 20°C (quite possible to have in a Lab)
  • Cold 15°C and Hot 40°C: average temperature is 27.5°C (quite impossible to have in a Lab and not so comfortable for researchers)
  • Cold 20°C and Hot 45°C: average temperature is 32.5°C (impossible to have in a Lab and not so comfortable for researchers)

Placing one or two TGR into the Conditioned Temperature Cabinet for TGR gives You the opportunity to set the appropriate temperature in order to have on the TGR a perfect temperature balance all along the ring and creating an optimal light and sound isolation ambient for the animals tested.

The cabinet is provided with an efficient air suction system to take out of the cabinet the hot air generated by two TGR machines and all the cable connections to drive from the cabinet outside the two TGR by a single ANY-Maze computer placed next to the cabinet.

Even if You can control the Lab temperature by a dedicated air conditioning system, the average temperatures needed for most of the TGR temperatures set-up makes the environment very uncomfortable for operators; by adopting the Conditioned Temperature Cabinet for TGR You can solve this obstacle and make Your tests in a perfect environment.

An extended warranty can be added for most Ugo Basile products. You can find this information in the ordering information section of each product.
To learn more about the extended warranty, CLICK HERE

Thermal Gradient Ring is based on thermal diffusion on a metal aluminum ring, cooled and heated on the two-extremity creating a 12 symmetrical temperature zones.

The environment where the machine is placed is critical in order guarantee the correct temperature balance and the correct thermal diffusion.

To achieve this goal the environment where the machine is installed needs to be at the average temperature between the two-extremity calculated with the following formula: (Cold Temp + Hot temp)/2

Typical temperature set-up are:

  • Cold 5°C and Hot 30°C: where average temperature result is 20°C
  • Cold 5°C and Hot 30°C: where average temperature is 27.5°C
  • Cold 20°C and Hot 45°C: where average temperature is 32.5°C

As You can imagine working in a Lab with 27.5 or 32.5°C is not comfortable and not under the labor safety rules; the use of the cabinet resolve this matter and You can place the cabinet inside the Lab whit out the need to control the Lab temp.

In addition to this You can control the lighting inside the TGR machine mixing IR and visible light to create a comfortable lighting environment for the animal, and last but not least, the animals inside the cabinet are sound isolated to eliminate external noise bias.

A complete set-up provides an efficient air suction unit (placed outside of the cabinet) to extract from the cabinet all the hot air generated by the cool/hot elements present in the TGR machine, to avoid temperature interaction with the cabinet temperature controller.

All the necessary cables are provided to power the one or two TGR and to connect the inside units to the ANY-Maze computer which is essential to run the experiment with one or two TGR units via a single computer.

ANY-maze TGR limited version

The ANY-maze TGR limited version gives You the ability to execute the TGR experiment only with a cheaper version of the software.

As an alternative, if You plan to use ANY-Maze for other video-tracking experiment You may choose to buy the ANY-Maze full version edition.

Remember that ANY-Maze is strictly linked to the PC where is installed and You can use one single ANY-Maze license at a time; if You plan to perform TGR and other behavioral experiment at the same time, You may need to buy 2 ANY-Maze license where You can have a multiple license discount.

TGR Measurement Parameters analyzed

  • Thermal zone occupancy: % of time spent in a zone.
  • Zone entries: number of entries in each zone
  • Preference temperature: calculated as the weighted preferred temperature
  • A “preference temperature” can be defined as the mean value using the zone temperature and “spent time” and the % of time spent above this temperature will be automatically calculated.
  • Cumulative distance: calculation of the cumulative distance per zone or per the whole apparatus
  • Coordinates of location within the Ring: Can be used to visualize mouse behaviour in the ring with Heat Maps
  • Safe, reproducible incubation, even at high ambient temperatures
  • Disinfection routine at +100 °C
  • Up to 30% lower energy consumption compared to the previous model
  • Hold one or two TGR units
  • Available for Eu and US power
  • Plug & Play system, all connection provided

General

Controls

Digital external from panel

Preset Temperature Read-out

Digital external from panel

Temperature range

from 0° up to 70° C (32° up to 158°F)

Temperature uniformity at 37°C

0.2 ± K

Temparature fluctuation at 37°C

0.1 ± K

Max heat compensation at 40 °C 500 W
Recovery time after door was opened for 30 s at 37 °C 3 min
Sound-pressure level 53 dB(A) (no air suction included)
Door lock lockable door handle

 

Electrical data

Rated Voltage 200...230 V
Power frequency 50/60 Hz
Nominal power 2.3 KW
Measures Interior Volume 698 L
Net weight of the unit (empty) 304 kg
Wall clearance back 100 mm
Wall clearance sidewise 100 mm

 

Physical

Width net 1261 mm
Height net 1925 mm
Depth net 1042 mm

 

Warranty

Warranty

2 years upon product registration on the Ugo Basile web site

Thermal conditioned cabinet for Thermal Gradient Ring

35680-EU

Thermal conditioned cabinet (for Europe 220/240 Volt EU plug). Temperature range from 0° up to 70° C (32° up to 158°F)

Back to: Pain and Inflammation

Research insights

Preclinical research in pain and inflammation

Pain affects an estimated 20% of adults worldwide, with millions of new chronic pain cases diagnosed every year. Yet, it remains insufficiently addressed as a major public health issue. Recognizing pain as a complex condition influenced by biological, social and environmental factors is essential to improve research, prevention, treatment and health policies.

Developing safer and more effective pain management strategies requires a deeper understanding of the biological mechanisms underlying pain and inflammation. In this context, preclinical research plays a fundamental role in identifying therapeutic targets, evaluating novel treatments and generating evidence to support translation into clinical practice.

Most used paradigms for pain and inflammation research

Pain and inflammation research uses a combination of evoked and spontaneous paradigms to assess nociception, hypersensitivity, inflammatory responses and pain-related impairment in rodents. Evoked thermal paradigms include plantar stimulation, tail flick, hot/cold plate testing, thermal place preference and thermal gradient assessment, which are used to investigate heat sensitivity, cold sensitivity, thermal hyperalgesia, allodinya and thermal preference or avoidance.

Mechanical paradigms include dynamic plantar stimulation, electronic or manual Von Frey testing, paw pressure testing and direct joint pressure application, supporting the assessment of mechanical allodynia, hyperalgesia and joint hypersensitivity. Spontaneous pain paradigms include weight-bearing asymmetry, postural imbalance, vertical activity and gait or locomotor changes, which are particularly relevant in arthritis, neuropathic pain, bone cancer pain, post-surgical pain and inflammatory models.

Inflammation can also be evaluated as a dedicated endpoint through paw volume and oedema measurement, allowing researchers to quantify experimentally induced swelling and pharmacological treatment effects.

Specific pain conditions like the ones involving the Trigeminal nerve are assessed by dedicated devices for thermal and mechanical stimulation of the animal vibrissae area, during the Orofacial Test.

Key behavioral readouts in pain assessment

Pain and inflammation studies rely on behavioural readouts that quantify nociceptive thresholds, hypersensitivity, inflammatory responses and spontaneous pain-related behaviour. Key measured parameters include paw withdrawal latency after thermal stimulation, tail flick latency, vocalization, jumping, mechanical withdrawal threshold, force threshold, paw volume, oedema, hind paw weight distribution, thermal preference or avoidance, time spent in specific temperature zones, postural asymmetry and pain-related changes in locomotion or vertical activity.

These readouts help researchers assess hyperalgesia, allodynia, inflammatory pain, joint pain, neuropathic pain, post-surgical pain and treatment effects across different rodent models. By combining evoked responses with spontaneous behavioural endpoints, preclinical pain research can capture both stimulus-induced sensitivity and more naturalistic pain-related impairment. Ugo Basile’s pain portfolio includes instruments for thermal, mechanical, inflammatory and spontaneous pain assessment, including Plantar Test, Dynamic Plantar Aesthesiometer, Hot/Cold Plate, Electronic Von Frey, Plethysmometer, Librae Incapacitance Tester and Climbing Test.

Ugo Basile instruments for pain and inflammation research

Since 1963, Ugo Basile has supported pain and inflammation research with instruments designed to measure nociceptive thresholds, hypersensitivity, inflammatory responses and spontaneous pain-related behaviour in rodents.

Ugo Basile offers a complete portfolio of instruments for preclinical pain and inflammation research, covering thermal, mechanical, inflammatory and spontaneous pain-related endpoints.

Ugo Basile devices for thermal stimulation

The Plantar Test is used for thermal hyperalgesia studies through radiant heat stimulation of the hind paw and automatic paw-withdrawal detection. It is the gold standard for Thermal Hyperalgesia tests in unrestrained animals. The Tail Flick measures nociceptive thresholds to infrared heat stimulation of the tail and is especially suited to rapid, reflex-based analgesic screening with possibility to run repeated tests thanks to the lack of learning being the tail flick a reflexive response. The Hot/Cold Plate supports hot and cold sensitivity studies, including constant temperature, ramping and complex temperature protocols, while the Thermal Place Preference and Thermal Gradient Ring allow researchers to evaluate thermal preference and avoidance in freely moving animals. Both systems involve a cognitive component and in fact repeated tests on the same animal are subject to hyper- or hypo- responses due to previous test experience.

Orofacial Stimulation measures hypersensitivity to thermal (hot and cold) and/or mechanical stimulation of the orofacial area in rats (mouse adaptor available) and is used for research into a wide range of pain conditions involving the trigeminal nerve with no need to cut vibrissae.

Ugo Basile devices for evoked mechanical stimulation

For mechanical sensitivity, the Dynamic Plantar Aesthesiometer provides automated force application and paw-withdrawal scoring for mechanical allodynia studies, while the Electronic Von Frey and Von Frey Filaments support mechanical threshold assessment across different force ranges. The PAM – Pressure Application Measurement system is designed for joint pain and arthritis models, enabling direct quantifiable stimulation of rodent joints. However the PAM can also be used with specific applicators for other parts of the body from the paws to the back and spinal chord, etc. Analgesy-Meter is typically used for inflammatory conditions to measure to measure mechanical sensitivity and pain threshold of pressure applied on mic or rat paw for basic or applied research or phenotyping.

Ugo Basile spontaneous pain apparatus

The Climbing Test adds a vertical activity endpoint for spontaneous pain, motor function and activity-related studies. Librae Incapacitance Tester assesses hind paw weight distribution as an objective measure of spontaneous pain. Plethysmometer measures paw volume and oedema.

Frequently Asked Questions

What instruments are used to measure pain in rodents?

Rodent pain can be assessed using different instruments depending on the type of pain response being studied. Thermal pain is commonly measured with plantar stimulation systems, tail flick meters, hot/cold plates, thermal place preference systems and thermal gradient paradigms. Mechanical sensitivity is assessed with dynamic plantar aesthesiometers, electronic Von Frey devices, manual Von Frey filaments, paw pressure systems and joint pressure tools. Spontaneous pain and inflammation can be evaluated through weight-bearing asymmetry, paw volume measurement, gait analysis and activity-based readouts.